Characterization Of Structures Of Equivalent Tissue With a Pixel Detector

M.C GRADOS LUYANDO* , B. DE CELIS ALONSO, E. MORENO BARBOSA, M.I. MARTÍNEZ HERNÁNDEZ, J.M. HERNÁNDEZ LÓPEZ AND G. TEJEDA MUÑOZ

1 Benemérita Universidad Autónoma de Puebla

*Email: carminagl87@gmail.com

Abstract 
Research using hybrid pixel detectors in medical physics is on the rise. Timepix detectors have arrays of 256 × 256 pixels with a resolution of 55 µm. Here, and by using Timepix counts instead of Hounsfield units, we present a calibration curve of a Timepix detector analog to those used for CT calibration. Experimentation consisted of the characterization of electron density in 10 different kinds of tissue equivalent samples from a CIRS 062M phantom (lung, 3 kinds of bones, fat, breast, muscle, water and air). Radiation of the detector was performed using an orthodontic X-ray machine at 70 KeV and .06 second of tube current with a purpose-built aluminum collimator. Data acquisition was performed at 1 frame per second and taking 3 frames per phantom. We were able to find a curve whose behavior was similar to others already published. This will lead to the verification of the usage of Timepix for identification of different tissues in an organ.